Measuring Principle
Ringbalance measuring system, factory-filled ringbody, supported by ball-bearings
Measuring Pickup
by magnetic field sensor, non-contact, with infinite resolution
Measuring Ranges
all ranges between 0-50 Pa (0.2" W.C.) to 0-1.8 kPa
(7" W.C.),
also plus/minus ranges such as -25/0/+25 Pa
(-0.1/0/+0.1"W.C.)
Scale units
Pa, daPa, kPa, mbar, mmWS, mmW.C.; in.W.C.
Accuracy
max. error +/-1.5% of span, below 100 Pa: +/-1.5 Pa
Outputs
0-20 mA, 4-20 mA und 0-10 V
Output Loads
mA - outputs: max. 680 Ohm, voltage - output: min. 2k Ohm
Power Supply
230 VAC (24/110/120/240 VAC; 24 VDC), 3.5 VA
Pressure Connections
2 fittings for flexible tubes, 8mm outside diameter
Electrical Connections
screw terminals, 2 cable fittings PG9 with strain relief
Housing
for wall mounting or panel mounting, made out of glass reinforced black polycarbonate, protection class IP40 (optional IP55)
Weight: 2.8 kg
Temperature
operating temperature 0...50°C (32...122 °F)
Electromagnetic Compatibility
according to EN 50081-2 (emission) and EN 50082-2 (immunity)
Option: EEx-i
The EEx-i version of the MU ANALOG consists of a Ringbalance-FP with integrated passive sensor, mounted in the hazardous area, in combination with the EEx-i Transducer type EXL-IMR10/pot, to be mounted in the safe area. Further information on request.
| Ordering specifications: | ||
| 1. | Type | MU ANALOG |
| 2. | Housing | wall- or panel mounting |
| 3. | Range | |
| 4. | Power supply | |
| ( 5. | Options ) | please specify |
| ( 6. | Overpressure ) | please note if you expect that the measuring range could be exceeded, (see: Overload-Protection ) |